MIL-DTL-6019E
shall then be applied and the rotation of the nut shall be limited by contact between a flat of the
insert nut and the case.
4.4.3.6 Mounting lugs. The type II gage case, with the mechanism and cover glass
removed, shall be so mounted in a suitable testing machine with the diameter of the case in a
horizontal plane that the mounting lugs receive no added support. A suitable pin shall be inserted
through the hole in the mounting nut and attached to a pull strap in the machine. A load of 175
pounds shall be applied to each lug for a period of 1 minute along the mounting hole axis and
toward the front of the case. There shall be no fracture.
5. PACKAGING
5.1 Packaging. For acquisition purposes, the packaging requirements shall be as
specified in the contact or order (see 6.2). When actual packaging of material is to be performed
by DoD personnel, these personnel need to contact the responsible packaging activity to ascertain
requisite packaging requirements. Packaging requirements are maintained by the inventory
control point's packaging activity within the military department or defense agency, or within the
military department's system command. Packaging data retrieval is available from the managing
military department's or defense agency's automated packaging files, CD-ROM products, or by
contacting the responsible packaging activity.
6. NOTES
(This section contains information of a general or explanatory nature that may be helpful,
but is not mandatory.)
6.1 Intended use. The pressure gage covered by this specification is intended for use in
indicating the pressure in cylinders installed in low-pressure oxygen systems in aircraft.
6.2 Acquisition requirements. Acquisition documents should specify the following:
a.
Title, number, and date of this specification.
b.
Type of gage (see 1.2).
c.
Specific issue of individual documents referenced (see 2.2.1 and 2.3).
d.
When a first article is required (see 3.1).
e.
Physical size and tolerance of cover glass (see 3.4.7.2).
f.
Required sampling plan tests A or B (see 4.4).
g.
Government specified sampling plan per MIL-STD-1916, if required (see 4.4.2 and 4.4.3).
h.
Packaging requirements (see 5.1).
6.3 Cross references. Parts with the above PINs are interchangeable as follows: