A-A-54519
Shall have an overall length of 152 millimeters (mm) +/- 5 mm. The top of the glass stem shall have a button top type finish that is perpendicular to the axial length of the thermometer. The button top diameter shall be greater than that of the body of the stem.
Workmanship. The thermometer shall be free from defects which detract from their appearance or impair their serviceability.
Unit. Each (EA). One thermometer, as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.
Inspections and tests. Inspections and tests shall be conducted in accordance with commercial practice to determine compliance with specification requirements. Sampling for inspections and tests shall also be conducted in accordance with commercial practice. In addition each thermometer shall be checked for graduation accuracy against standard thermometer(s) whose calibration is traceable to the National Institute of Standards and Technology (NIST). The graduation markings selected for testing shall include the +4 and +6 degree (C) temperature divisions. A thermometer that does not conform to the +/- 1 degree (C) accuracy requirement at any point on the required scale marking shall be cause for rejection.