A-A-53034
In operation, the thermometer shall be instant recording and the time for a thermometer reading to rise from 50 to 160 degrees (F) shall not exceed 10 seconds (see Quality Assurance Provisions).
The thermometer shall be provided with printed instructions for use, adjustment and care of the thermometer.
Workmanship. The thermometer shall be free from defects which detract from its appearance or impair its serviceability.
Unit. Each (EA). One thermometer, as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.
Tests. Tests shall be conducted to determine compliance with specification requirements. Where feasible, the same sample shall be used for the determination of two or more test characteristics.
Accuracy testing. The thermometer shall be tested against a standard thermometer at 32 degrees (F), and at two additional temperature points so that there is at least a 50 degree difference between the middle and the other two readings. Sampling for testing shall be in accordance with MIL-STD-105 and an S-3 inspection level. The acceptance number shall be zero. Any thermometer reading that is not within ± 2 degrees of the standard thermometer reading shall be cause for rejection.